1. Class I Biosafety Cabinet does not protect the product from contamination because un-purified
room air constantly enters into work area.
2. As a partical containment unit, the Class I Biosafety Cabinet is suitable for work involving low to
moderate risk agents (Biosafety Levels 1, 2 and 3) where there is a need for containment, but not for
product protection.
3. Unlike conventional fume hood, the HEPA filter in the Class I Biosafety Cabinet protects the
environment by filtering air before it is exhausted.
4. With the negative pressure, personnel protection is made possible by constant movement of air
into the work area.
Features:
- Soft touch control panel, LCD display.
- UV lamp for sterilization.
- Stable air flow system; protecting to person and environment.
- Energy-saving, high efficiency, low noise.
- One piece 304 stainless steel work stable, easy for cleaning.